HFA3 Instructions for Use 2660021166131 Rev. A 2018-11
Go to Contents Specifications
12-1
(12) Specifications
HFA3 Instrument
Stimulus
Maximum Intensity: 10,000 ASB
Duration: 200 msec, 500 msec for Esterman tests, continuous for kinetic tests
Wave Length: Broadband visible light
Visual Field Testing Distance
30 cm
Bowl Illumination
•31.5 ASB
Maximum Temporal Range
90 degrees (90°)
Dynamic Range
50 dB
Computer
Operating System: Windows 7 or 10 – 64 bit
•Intel
®
Core i7-610E
Internal storage: 500 GB
6 external USB ports, Type A, USB 2.0 specification
Ethernet Port (4KV isolated GbE)
DisplayPort video output
•12V power outlet
•USB Keyboard & Mouse
LCD Display with integrated touch screen
Physical
Dimensions: 46 L x 52 W x 58 H (cm)
Weight: 28.7 kg (63 lbs.)
Electrical Requirements
Electrical Rating
•100120V~, 50/60Hz, 4.0A
230V~, 50/60Hz, 1.8A
Environmental Conditions
Transport (Packaged)
Temperature: –40 to +70°C
Humidity: 10% to 95%
Atmospheric Pressure: 500 to 1060 hPa
Vibration, sinusoidal: 10 Hz to 500 Hz: 0,5 g
Shock: 30 g, duration 6 ms
Bump: 10g, duration 6 ms

Table of Contents

Other manuals for Zeiss Humphrey Field Analyzer 3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Zeiss Humphrey Field Analyzer 3 and is the answer not in the manual?

Zeiss Humphrey Field Analyzer 3 Specifications

General IconGeneral
BrandZeiss
ModelHumphrey Field Analyzer 3
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals