HFA3 Instructions for Use 2660021166131 Rev. A 2018-11
Go to Contents Setup and Testing
4-13
Note: The patient can pause the test by holding down the response button, which will result in an
on-screen message and an audible alert. The Start of
Test sound will occur when the patient
releases the button and the test is resumed.
RelEYE (Models 850 and 860)
Turn RelEYE on to record images of the patient’s eye during the presentation of each stimulus (see
“Specific Settings,” on page 3-13). This allows the operator to determine how well
the patient was
fixating at a specific test point.
Eye images will be transferred along with the patient
exam data to FORUM. When the instrument is
in local database mode RelEye images can only be viewed during and directly at the end of a test,
prior to selecting Review Results (see page “Review and Save Results,” on page 4-14).
To review RelEye images:
1. Select a tested point on the test pattern. Already tested points are
displayed
as shaded circles. An icon will appear centered over the
point, along with associated, multiple eye images to the left.
2. Drag the targeting icon over different test points to view the
r
ecorded eye images. Touch or click the
icon to return to a
view of the live eye.
RelEYE is available for the following tests only:
• Threshold SITA-Standard tests (Central 30-2, Cen
tral 24-2, Central 10-2, and Peripheral 60-4
test patterns)
• Suprathreshold Central 76
• Suprathreshold Esterman Monocular