Guidance and manufacturer’s declaration - electromagnetic emissions
The HFA3 is intended for use in the electromagnetic environment specified below
. The customer or user of the HFA3 should ensure that it is
used in such an environment.
Emissions Test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1 The HFA3 uses RF energy only for its internal
function. Ther
efore, its RF emissions are very
low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A The HFA3 is suitable for use in all
establishmen
ts other than domestic
establishments and those connected to a low
voltage power supply network which supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Complies
Guidance and manufacturer’s declaration - electromagnetic immunity
The HFA3 is intended for use in the electromagnetic environment specified below
. The customer or user of the HFA3 should ensure that it is
used in such an environment
Immunity Test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Electrostatic Discharge
(ESD
) IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete, or ceramic tile. If
floors are covered with synthetic material, the
relative humidity should be at least 30%.
Electrical fast
tr
ansient/burst
IEC 61000-4-4
± 2 kV for power supply
lines
± 1 kV for input/output
lines
± 2 kV for power supply
lines
± 1 kV for input/output
lines
Mains power quality should be that of a typical
commer
cial or hospital environment.
Surge IEC 61000-4-5 ± 1 kV differential mode
± 2 kV common mode
± 1 kV differential mode
± 2 kV common mode
Mains power quality should be that of a typical
commer
cial or hospital environment.
Voltage dips, short
in
terruptions, and voltage
variations on power supply
input lines. IEC 61000-4-11
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
) for
5 cycles
70% U
T
(30% dip in U
T
) for
25 cycles
<5% U
T
(95% dip in U
T
)
for 5 sec.
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
) for
5 cycles
70% U
T
(30% dip in U
T
) for
25 cycles
<5% U
T
(95% dip in U
T
)
for 5 sec.
Mains power quality should be that of a typical
commer
cial or hospital environment. If the user of
the HFA3 requires continued operation during
power mains interruptions, it is recommended that
the HFA3 be powered from an uninterruptible
source.
Power Frequency
(50/60 Hz) magnetic field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic fields should be at levels
char
acteristic of a typical location in a typical
commercial or hospital environment.
Note: U
T
is the a.c. mains voltage prior to application of the test level.
Introduction Go to Contents
HFA3 Instructions for Use 2660021166131 Rev. A 2018-11
1-10

Table of Contents

Other manuals for Zeiss Humphrey Field Analyzer 3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Zeiss Humphrey Field Analyzer 3 and is the answer not in the manual?

Zeiss Humphrey Field Analyzer 3 Specifications

General IconGeneral
BrandZeiss
ModelHumphrey Field Analyzer 3
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals