EasyManua.ls Logo

ABB SCK - Page 119

ABB SCK
244 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
OI/SCK
-EN Rev. E Operating instruction SCK Gas Sampling System 119
Proceed as follows to switch on the gas sampling system:
Step
Procedure
1
Shut off the compressed-air using the gate valve
(1)
in the
pneumatic box of the probe retractor so as to prevent any
undesirable probe movements.
2
Completely vent the compressed-air tank (if present).
3
Open the control cabinet.
4
Turn on the main switch of the gas sampling system.
5
Switch on uninterruptible power supply (if present).
6
Switch on residual current circuit-breaker in the control cabinet
(if present).
7
Flip the fuse switches one by one in the control cabinet.
8
Wait for controller to boot.
9
Start frequency converter.
10
Adjust local time (password for "Advance Service" required).
Path: Setup System Time setting
11
Open the compressed-air gate valve
(1)
again.
WARNING! Depending on the current position, the probe be-
gins to move towards to the rear end position (probe com-
pletely retracted). Do not stand in the area of travel of the
probe or probe carriage.
12
Switch on the cooling-water pump on the touchscreen.
Path: Operation Cooling Water Pump ON
13
Wait for the heated modules to heat up until the following error
messages disappear:
"Probe heating temperature MIN"
"Cable Drag Chain Sample Gas Line Temp. Error"
Path: Diagnosis Message
14
Turn service switch to
ON
position.
Note: This step resets various safety functions.
15
Turn service switch to
OFF
position.
Result: The gas sampling system is in manual mode.
Switching on the gas
sampling system.

Table of Contents

Other manuals for ABB SCK

Related product manuals