118 Operating instruction SCK Gas Sampling System OI/SCK-EN Rev. E
Activation
Risk of injury in the area of the gas sampling system due to moving or fall-
ing parts, hot surfaces, hot cooling water, anti-freeze, compressed air,
lubricating oil
Wear general personal protective equipment:
Closed overalls with long trousers and long sleeves
Safety shoes
Safety gloves suitable for mechanical work
Safety glasses
Protective helmet
If necessary, additional protective equipment if prescribed by the opera-
tor.
Danger of electric shock when control cabinet is open, open connection or
terminal boxes
Only specialized electrical personnel may work on the electrical system.
Observe national regulations pertaining to electrical equipment.
The main terminals -X30, -X40 and the input terminals of the main
switches -Q30, -Q40 are live even when the main switches are off.
All metallic parts in the environment that can be touched at the same
time as the machine must also be grounded together.
Check pre-protection according to the documentation.
Risk of crushing around the retractor due to gas sampling probe move-
ments
Do not stand in the danger area around the retractor.
Press the EMERGENCY-STOP switch immediately if anyone is still in the
danger area around the retractor after the warning lamps light up or the
warning buzzer buzzes.
The connected analyzer system must be switched on. Observe the warm-
up time of the analyzer system.
The mechanical locking bar of the shutter must be removed before insert-
ing the gas sampling probe for the first time.
The system must not have any visible damage (see also
Checking general
condition
on page 141).
The customer-side central compressed-air supply is switched on.
Control cabinet key to open the control cabinet
Mechanical, thermal
and material dangers
components
movements
Requirements