EasyManua.ls Logo

ABB SCK - Page 125

ABB SCK
244 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
OI/SCK
-EN Rev. E Operating instruction SCK Gas Sampling System 125
Proceed as follows to start the frequency converter and adapt the control
parameters:
Step
Procedure
1
Open the control cabinet.
2
In the control cabinet: Start the frequency converter ACS 355
(1)
by pressing the START button.
Check whether "REM" (for remote mode) is indicated on the left
side of the display and change by pressing the LOC REM button
if required.
Result: The frequency converter displays "A5011" for approx.
2 seconds.
Note: The frequency converter must be restarted after each
system power failure. The following steps 3 to 10 must be
executed only during the first system start-up and not after a
system power failure.
3
Close control cabinet.
4
Turn service switch to OFF position.
5
Insert probe.
6
Monitor cooling water return temperature.
Path: Diagnosis Trends Cooling water return
Note: If the set points are exceeded due to inadequate adapta-
tion of the control parameters, the program sequences for the
protection of the probe are started automatically.
7
Adjust control parameter KP in small steps until the cooling
water return temperature runs as constant as possible (de-
fault: -2).
Path: Setup FC parameters
Note: Enter the KP value always negative using the +/button.
8
Adjust control parameter KI to further optimize the tempera-
ture run (TN, default: 10).
9
If necessary, also adjust control parameter KD (TV, default: 0).
10
Save control parameter settings using the button.
Instructions

Table of Contents

Other manuals for ABB SCK

Related product manuals