82 Operating instruction SCK Gas Sampling System OI/SCK-EN Rev. E
Danger Where / in what situa-
tion does the danger
arise?
Preventive action
Danger due to
current-carrying
parts
When working on elec-
tric components
Only specialized electrical
personnel may work on
the electrical system.
Disconnect the gas sam-
pling system from the
power supply before
working on other electric
components.
Observe national regula-
tions pertaining to elec-
trical equipment.
The main terminals -X30,
-X40 and the input termi-
nals of the main switches
-Q30, -Q40 are live even
when the main switches
are off.
All metallic parts in the
environment that can be
touched at the same time
as the machine must also
be grounded together.
Check pre-protection
according to the docu-
mentation.
Risk of burning due
to hot surfaces
The gas sampling
system is very hot
during and after opera-
tion of the gas sam-
pling system, particu-
larly:
the cooling module
the heated sample
gas lines
the connection box
for the heated sam-
ple gas line on the
retractor
the heated sam-
pling filter
the entire gas sam-
pling probe
the metal structure
of the retractor
the shutter of the
duct opening
Do not touch the gas
sampling system during
and after operation.
Do not touch the gas
sampling system until it
has cooled down to less
than 50 °C.
If it is unavoidable to
touch during operation,
wear a face mask against
heat, heat-proof gloves
and protective welder's
clothing.