EasyManua.ls Logo

ABB SensyMaster FMT430 - Page 121

ABB SensyMaster FMT430
160 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SensyMaster FMT430, FMT450 THERMAL MASS FLOWMETER | OI/FMT430/450-EN REV. B 121
Menu / parameter
Description
Communication / ...Profibus
Address
Set the PROFIBUS DP® device address (1 to 126).
Ident Nr. Selector
Display the
PROFIBUS DP® identification number
0x9740 -PA 1AI+1TOT
0x3435 -ID Specific
Adaptation Mode
0x9700 -PA 1AI
Comm State
Display the PROFIBUS communication status.
Offline: No PROFIBUS® communication.
Stop: Bus active, device not active.
Clear: Device is being initialized.
Operate: Cyclic communication is active.
Baud Rate
Display the transmission speed (baud rate) for the PROFIBUS® communication.
The baud rate is automatically detected and does not need to be configured manually.
PB
Manufacturer ID
Display the PROFIBUS
DP® manufacturer ID
26: ABB

Table of Contents

Related product manuals