SensyMaster FMT430, FMT450 THERMAL MASS FLOWMETER | OI/FMT430/450-EN REV. B 121
Menu / parameter
Description
Communication / ...Profibus
Address
Set the PROFIBUS DP® device address (1 to 126).
Ident Nr. Selector
Display the
PROFIBUS DP® identification number
•
0x9740 -PA 1AI+1TOT
•
0x3435 -ID Specific
•
Adaptation Mode
• 0x9700 -PA 1AI
Comm State
Display the PROFIBUS communication status.
•
Offline: No PROFIBUS® communication.
•
Stop: Bus active, device not active.
•
Clear: Device is being initialized.
• Operate: Cyclic communication is active.
Baud Rate
Display the transmission speed (baud rate) for the PROFIBUS® communication.
The baud rate is automatically detected and does not need to be configured manually.
PB
Manufacturer ID
Display the PROFIBUS
DP® manufacturer ID
• 26: ABB