SensyMaster FMT430, FMT450 THERMAL MASS FLOWMETER | OI/FMT430/450-EN REV. B 131
Error messages
Error number
Error text
Description
NAMUR
classification
F098.011
No Frontend Board detected.
Wrong connection. Defect
Frontend.
Communication error to frontend board (FEB) of the sensor.
No
frontend board found, frontend board defective.
For remote mount design: check signal cable connection between sensor and
transmitter.
Restart the device
Replace frontend board
Contact ABB Service
Failure
F096.029
ADC Failure on Frontend
Board.
AD
-converter error in frontend board (FEB) of the sensor.
EMC defects, frontend board defective
Check installation
Restart the device
Replace frontend board
Contact ABB Service
Failure
F094.021
Safety Alarm Curr.Out 31/32
Error in
current output 31 / 32 / Uco
Check electrical connection current output 31 / 32 / Uco
Contact ABB Service
Failure
F093.033
Sensor failure or
disconnected.
Electrical connection of the sensor or the thermal measuring element is defective. Sensor
not connected, signal cable defective, sensor element defective.
For remote mount design: check signal cable connection between sensor and
transmitter.
Check thermal measuring element
Contact ABB Service
Failure
F092.030
Electronics
failFrontend
Board.
Electronics error in frontend board (FEB) of the sensor.
EMC defects, frontend board defective
Check installation
Restart the device
Replace frontend board
Contact ABB Service
Failure
F088.012
FEB communication error.
Communic
ation error to frontend board (FEB) of the sensor.
Electro magnetic interference.
For remote mount design: check signal cable connection between sensor and
transmitter.
Restart the device
Contact ABB Service
Failure
F086.018
Curr.Out 31/32 com
error.
Communication error to current output 31
/ 32 / Uco.
Electro magnetic interference, faulty motherboard in transmitter.
Contact ABB Service
Failure
F084.010
NV data defect. Data storage
irreparable.
Error in SensorMemory.
Memory module defective
Contact ABB Service
Failure
F082.013
Incompatible Frontend Board.
Incompatible frontend board.
The frontend board is not compatible with the motherboard in the transmitter.
Contact ABB Service
Failure
F081.041
FEB voltages outside range.
Frontend board power supply defective. Frontend board defective.
Check power supply
Replace frontend board
Contact ABB Service
Failure
F080.025
MB voltages outside range.
Motherboard power supply defective Faulty motherboard.
Contact ABB Service.
Failure