SensyMaster FMT430, FMT450 THERMAL MASS FLOWMETER | OI/FMT430/450-EN REV. B 21
Changing the type of protection
If you are installing in Zone 1 / Div. 1, the current outputs and digital outputs of models FMT430/450 can be operated with different
types of protection:
Current output and digital output in the ‘intrinsically safe ia / IS’ design
Current output and digital output in non-intrinsically safe design
If a device that is already operational is operated with a different type of protection, the following measures must be
implemented/insulation checks performed in accordance with applicable standards.
Original installation
New installation
Necessary test steps
Zone 1 / Div.
1:
Current outputs and digital outputs in
non
-intrinsically safe design
Zone 1 / Div.
1:
Current outputs and digital outputs in
intrinsically safe ia
/ IS design
500 × 1.414 = 710 V DC/1min
Test between terminals A / B, U
FE
, /GND, U
CO
/ 32, 31 / 32, 41 / 42, 51 / 52,
V1 / V2 and V3 / V4, and terminals A, B, U
FE
, GND, U
CO
, 31, 32, 41, 42, 51, 52,
V1, V2, V3, V4 and the housing.
When this test is performed, no voltage flashover is permitted in or on the
device.
Optical evaluation particularly of the electronic circuit boards, no visible
damage or evidence of explosion.
Zone 1 / Div.
1:
Current outputs and digital outputs in
intrinsically safe ia(ib) / IS design
Zone 1 / Div.
1:
Current outputs and digital outputs in
non-intrinsically safe design
Visual inspection, no damage visible on the threads (cover, ½ in NPT cable
glands).