SensyMaster FMT430, FMT450 THERMAL MASS FLOWMETER | OI/FMT430/450-EN REV. B 39
Mounting
Calculation of mounting dimensions
Length of the welding adapter
Height dependent on nominal
diameter
h3 Installation length of the sensor
∅D Outside diameter of the piping
Flow direction (arrow marker on the protection tube)
Compression fitting
Safety snap ring
Figure 24: Calculating mounting dimensions
Calculations (mm)
L = h3−(½×ØD)
Z = (h3+137 mm)−(½×ØD)
Calculations (in)
L = h3−(½×ØD)
Z = (h3+5.39 in)−(½×ØD)
Preparing the sensor
Fire hazard in oxygen applications
Fire hazard in oxygen applications due to the use of
unapproved thread sealing compound.
• Use only approved thread sealing compound for oxygen
Risk of injury
Risk of injury due to the sensor ejecting because of a missing
safety ring.
• Mount the sensor with compression fitting only with the
Sensor protection tube
Safety snap ring
Marking for reinstallation
Figure 25: Snap ring and marking
1. Slide the compression fitting onto the sensor and tighten by
hand so that the compression fitting can still be moved.
2. Insert the safety snap ring using mounting pliers (see
Figure 24, Pos.
2).
Note
For gas-tight sealing of the NPT thread of the compression
fitting, you can for example use special thread sealing
compounds by Swagelok such as SWAK™, Silver Goop™, PTFE-
Free, etc., or PTFE thread sealing tape.