SensyMaster FMT430, FMT450 THERMAL MASS FLOWMETER | OI/FMT430/450-EN REV. B 125
Menu: Totalizer
...Operation’ using .
Selection of submenu ‘...Reset Totalizer’ using .
...Preset Totalizer’ using .
Selection of submenu ‘...FillMass’ using .
counters.
...Reset Totalizer
Resets all totalizers to zero.
Reset all mass totalizers to zero.
Reset all standard volume totalizers to zero.
Totalizer / ...Preset Totalizer
Input from meter readings (e.g. when replacing the transmitter).
...FillMass
Selection of process variable used during the filling process.
Off: Filler deactivated.
VolumeFlow@: Standard flow rate.
Sets the fill quantity using the selected unit.
When the defined fill quantity is reached, the configured binary output is activated.
fill quantity, the corresponding process value must be selected with the parameter ‘Batch Process Value’.
Resets the current fill quantity.
Manual start of the filling function.
Alternatively, the digital input can
be configured for starting / stopping the fill operation.
Display of the current fill quantity.
Once a fill operation has been started, the quantity already filled is shown here. The counter restarts at zero for each fill
operation initiated and then counts up to the set fill quantity.
Manual stop of the filling function.
Alternatively, the digital input can be configured for starting / stopping the fill operation.
Display of the number of fill operations since the last reset.
Current Batch Counts’ to zero.
Select the ‘...Quantity’ submenu.