SensyMaster FMT430, FMT450 THERMAL MASS FLOWMETER | OI/FMT430/450-EN REV. B 133
Error number
Error text
Description
NAMUR
classification
S045.034
Sensor heat emission limit.
Heat emission limits of the measuring element
up-scaled. Flow rate too high, incorrect
measuring medium
Check the process conditions
Out of specification
S044.000
Mass flowrate exceeds limits.
The mass flow is below or above the configured limit values ‘
Qm Massflow Min’ and ‘Qm
Massflow Max
’.
Check the settings of the parameters in the ‘Process Alarm / ...Alarm Limits’ menu and
adjust if necessary.
Check mass flow.
Out of specification
S042.037
Medium temperat exceeds
limits.
The measuring medium temperature is below or above the configured
limit values ‘Tm
Temperature Min
’ and ‘Tm Temperature Max’.
Check the settings of the parameters in the ‘Process Alarm / ...Alarm Limits’ menu and
adjust if necessary.
Check measuring medium temperature.
Out of specification
S041.039
Std.Volume flow exceeds
limits.
The standard volume flow is below or above the parameterized limit values "
Qv@
Volumeflow Min
" and "Qv@ Volumeflow Max".
Check the settings of the parameters in the ‘Process Alarm / ...Alarm Limits’ menu and
adjust if necessary.
Check volume flow rate.
Out of specification
M059.038
Invalid Sensor configuration
Parameterization (configuration) of the device is incorrect.
Check parameterization (configuration)
Contact ABB Service
Maintenance
required
M058.040
Sensor soiling detected.
Thermal measuring element contaminated.
Check thermal sensor elements and clean as needed Maintenance on page 135)
Maintenance
required
M038.009
Sensor memory defective.
SensorMemory missing or defective
Check SensorMemory
Replace SensorMemory
Contact ABB Service
Maintenance
required
M037.014
NV chips defect on
Motherboard.
Memory module on the motherboard defective.
Contact ABB Service
Maintenance
required
M032.022
Curr.Out 31/32 not calibrated.
Current output 31
/ 32, Uco not calibrated.
Contact ABB Service.
Maintenance
required
M031.023
Curr.Out V1/V2 not calibrated.
Current output (plug
-in card) V1 / V2 or V3 / V4 not calibrated.
Check plug-in card and replace if necessary.
Contact ABB Service.
Maintenance
required
M030.024
Curr.Out
V3/V4 not
calibrated.
M028.007
Display value is<1600h at
Qmax.
The current counter reading has exceeded the display resolution.
Check the unit setting for the mass counters / volume totalizers and adjust if
necessary.
Maintenance
required
M026.004
Maintenance interval is
reached
Maintenance interval reached.
Perform maintenance work.
Start new maintenance interval in the ‘Diagnostics / ...Diagnosis Control’ menu.
Maintenance
required
M024.008
Device not calibrated.
Have device
calibrated.
Contact ABB Service.
Maintenance
required
M020.027
Communicat. card not
responding
Fieldbus plug
-in is not reacting.
Plug-in card is defective.
Contact ABB Service.
Maintenance
required