System operation 33
fabian Therapy evolution | SW 5.1.x
Ref: 121003.EN / Date 2020-01-22
5.5 System shutdown
Switching OFF the fabian
1. Remove patient from the ventilator.
2. Power OFF and disconnect any
SpO
2
devices connected to the
ventilator.
3. Press and hold the ON
∕ OFF button (1) for 6
seconds.
4. After 6 seconds, a menu will appear
on the screen providing an option to
shut down the device.
5. Press Shut down to power OFF the
device.
6. After the device has powered OFF, a
notice tone will sound.
7. Press the Alarm Silence button to confirm system switch OFF.
8. Hold the button for at least three seconds until the WARNING
LED stops.
5.6 Emergency shutdown
WARNING
In the event of a fault resulting in loss of normal control the emergency shutdown
procedure may be used . This procedure disconnects all power to the ventilator and any
other devices connected to the ventilator.
Before performing this procedure, ensure that the alternate ventilation method (for
example: manual resuscitation) is present, ready and standing by.
1. Remove patient from the ventilator.
2. Press and hold the ON ∕ OFF button
(1) for one second.