Alarms 51
fabian Therapy evolution | SW 5.1.x
Ref: 121003.EN / Date 2020-01-22
7.1.1.2 Automatic alarm presets
The following alarm limit autosets are used as defaults.
Breaths per Minute (BPM) upper limit: 150% of measured breath rate
APNEA: 10 seconds
Pduo [DUOPAP-mode] upper limit: 5 mbar above set P
duo
value
Pduo [DUOPAP-mode] lower limit: 2 mbar above PEEP lower limit
CPAP [DUOPAP-mode] lower limit: 2 mbar below set CPAP value
CPAP [NCPAP-mode] upper limit: 5 mbar above set CPAP value
CPAP [NCPAP-mode] lower limit: 2 mbar below set CPAP value
7.1.2 Configurable alarms
Following alarm limits are displayed and can be configured in the
Alarm Limits menu.
High SpO
2
alarm: 2 to 99%, off
Low SpO
2
alarm: off, 1 to 98%
High Pulse Rate alarm: 35 to 235 bpm, off
Low Pulse Rate alarm: off, 30 to 230 bpm
Low Perfusion Index alarm: off, 0.03 to 19%
Low SIQ Alarm: off, 5 to 100%.