8 Battery operation
The device features an Internal Battery.
Battery run time in DUOPAP mode on full charge is approximately 3 hours
These values may vary depending on the parameter settings. The battery charge is continusously
monitored. The device continuously calculates the remaining Run Time and displays it in the
Information bar.
WARNING: A reduction in Battery remaining run time when changing the
Operating mode can cause an Automatic shutdown of the ventilator!
Changing the Operating mode can greatly shorten the battery run time.
• Monitor the remaining Battery run time when changing the settings.
• NEVER leave the device and patient unattended during battery
operation.
• Establish mains power supply in good time!
In the event of Power Failure, the pneumatic system will automatically open the
safety valve to
breathing system, allowing spontaneous patient breathing.
When storing the ventilator for long term storage, please remove the Battery
pack.
In the event of a Power Failure, the device
will automatically switch to Battery power
without interruption.
A visual and audible message will be
triggered.
Power failure
This message can be
confirmed with the Alarm
Silence button.
The device will continuously monitor the battery run time.
Additional messages will be triggered at a remaining Run Time of:
• 15 minutes
• 30 minutes
• 60 minutes, only if the remaining run time is less than 60 minutes during a Power
Failure