EasyManua.ls Logo

ACUTRONIC fabian Therapy evolution - PRICO Disabling Alarms

ACUTRONIC fabian Therapy evolution
106 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
76 Accessories and options
fabian Therapy evolution | SW 5.1.x
Ref: 121003.EN / Date 2020-01-22

2
Flushan be activated during PRICO, this will deactivate PRICO.
To re-enable PRICO after an O
2
Flush
PRICO menu.
To disable the O
2
Flush function, set the Flush Time to 0s
 Stop button could be used to interrupt mechanical ventilation for 2
minutes.
In nCPAP, DUOPAP and O
2
Therapy mode ventilation can be interrupted
indefinitely.
In both cases PRICO will be switched OFF.
When PRICO is ON, this is indicated by the PRICO icon in the Status bar and the green field with the
current FiO
2
labelled PRICO above the O
2
parameter.
PRICO 
12.3.4 PRICO disabling alarms
PRICO is turned OFF and cannot be re-enabled while any of the following alarms are active:
Air supply pressure
O
2
value out of range
Check ET tube
Oxygen sensor defect
Clean flow sensor
Oxygen sensor used up
Check SpO
2
sensor
Oxygen supply pressure
Flow sensor defect
Patient disconnected
Flow sensor not connected
SpO
2
module disconnected

Table of Contents

Related product manuals