76 Accessories and options
fabian Therapy evolution | SW 5.1.x
Ref: 121003.EN / Date 2020-01-22

2
Flushan be activated during PRICO, this will deactivate PRICO.
• To re-enable PRICO after an O
2
Flush
PRICO menu.
• To disable the O
2
Flush function, set the Flush Time to 0s
 Stop button could be used to interrupt mechanical ventilation for 2
minutes.
In nCPAP, DUOPAP and O
2
Therapy mode ventilation can be interrupted
indefinitely.
In both cases PRICO will be switched OFF.
When PRICO is ON, this is indicated by the PRICO icon in the Status bar and the green field with the
current FiO
2
labelled PRICO above the O
2
parameter.
PRICO 
12.3.4 PRICO disabling alarms
PRICO is turned OFF and cannot be re-enabled while any of the following alarms are active:
• Air supply pressure
• O
2
value out of range
• Check ET tube
• Oxygen sensor defect
• Clean flow sensor
• Oxygen sensor used up
• Check SpO
2
sensor
• Oxygen supply pressure
• Flow sensor defect
• Patient disconnected
• Flow sensor not connected
• SpO
2
module disconnected