Electromagnetic compatibility statement 99
fabian Therapy evolution | SW 5.1.x
Ref: 121003.EN / Date 2020-01-22
18.2 Device with SN till AT-04000
Guideline and manufacturer declaration – electromagnetic emission
The device "fabian Therapy evolution" is intended for operation in the environment described below. The
customer or user of the "fabian Therapy evolution" apparatus should ensure it is operated in this type of
environment.
Electromagnetic environment – guideline
The device "fabian Therapy evolution" uses HF energy
solely for internal operation. Its HF emission is thus very
low and interference with adjacent apparatuses is unlikely.
The device fabian Therapy evolution is suitable for use in
all other areas except residential areas and areas directly
connected to a public mains supply which also supplies
buildings used for residential purposes.
Emission of harmonics
IEC 61000-3-2
Emission of voltage
fluctuations / flickers
IEC 61000-3-3
Guideline and manufacturer declaration – electromagnetic immunity
The device fabian Therapy evolution is intended for operation in the environment described below. The
customer or user of the fabian Therapy evolution apparatus should ensure it is operated in this type of
environment.
Electromagnetic environment –
guideline
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wooden or
concrete or tiled with ceramic tiles.
For flooring made from synthetic
materials the relative air humidity
must not be less than 30%.
Quick electric
transients/burst
IEC 610004-4
± 2kV for power
cords
± 1kV for in-
/output cables
± 2kV for power cords
± 1kV for in-/output
cables
The mains quality should
correspond to typical business or
hospital environments.
± 1 kV lead to
lead
± 2 kV lead to
ground
± 1 kV lead to lead
± 2 kV lead to ground
The mains quality should
correspond to typical business or
hospital environments.