Alarms 53
fabian Therapy evolution | SW 5.1.x
Ref: 121003.EN / Date 2020-01-22
7.4 Alarm causes and solutions
Alarms are categories by three main priority levels; High, Medium and Low. They differ visually and
acoustically according to priority.
I=HIGH blinking message highlighted Red --- 4 second pause
II=MEDIUM blinking message highlighted Yellow  5 second pause
III=LOW message highlighted Yellow  15 second pause
Only one alarm can be signalled at a time. In case there is more than one alarm condition at the
same time, only the highest priority alarm condition is active and displayed. In the event there are two
alarm conditions with the same priority at the same time, the first alarm condition triggered is active
and displayed.
7.4.1 Alarms table
This table is intended to help you determine and resolve the cause of an alarm message.
#
Alarm Text
Alarm Type
Cause
Solution
Main
Priority
Sub
Priority
1.
Acoustic audio
alarm test
failed
System Failure
Acoustic audio test has
failed
Disconnect the device
from AC power and
contact your local
Acutronic Medical
Systems Distributor
I
1
2.
Cooling fan
defect
System failure
Fan not moving
Disconnect the device
from AC power and
contact your local
Acutronic Medical
Systems Distributor
I
1
3.
COM interface
System failure
System error
Disconnect the device
from AC power and
contact your local
Acutronic Medical
Systems Distributor
I
1
4.
DIO interface
System failure
System error
Disconnect the device
from AC power and
contact your local
Acutronic Medical
Systems Distributor
I
1
5.
I2C interface
System failure
System error
Disconnect the device
from AC power and
contact your local
Acutronic Medical
Systems Distributor
I
1
6.
parallel
interface
System failure
System error
Disconnect the device
from AC power and
contact your local
Acutronic Medical
Systems Distributor
I
1