EasyManuals Logo

ACUTRONIC fabian Therapy evolution User Manual

ACUTRONIC fabian Therapy evolution
106 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #65 background imageLoading...
Page #65 background image
Ventilation modes 65
fabian Therapy evolution | SW 5.1.x
Ref: 121003.EN / Date 2020-01-22
11 Ventilation modes
The ventilator is intended for the following Ventilation modes:
nCPAP : NIV ventilation (Optional)
duoPAP : NIV ventilation (Optional)
O
2
Therapy: High and Low Flow Oxygen Therapy HFNC (Optional)
11.1 nCPAP DUOPAP
In nCPAP DUOPAP mode the patient breathes spontaneously from a mask or nasal prongs.
A NIV Trigger option can be enabled to provide these modes with breath detection (Apnea
monitoring with Alarms) and triggered supported breaths (or Synchronized Transition from low to
high CPAP levels) in DUOPAP.
Depending on the nasal system generator being used, fabian will then automatically select the proper
NIV Trigger Sensor:
Flow Sensor (Neonatal) with Infant Flow (Infant Flow or Inspire) and Infant FlowLP generators
Pressure Sensor with Medijet (by Medin) generator
Additionally, monitoring SpO
2
and PCO
2
is always recommended.
nCPAP: Supplies a positive Airway Pressure with automatic Leak Compensation.
The maximum flow compensation is selectable by menu.
DUOPAP: Same as nCPAP but with the option of Positive Pressure Ventilation with
adjustable frequency and Inspiratory Period.
This mode requires a special nCPAP Patient Set with nCPAP generator.
Before using the nCPAP DUOPAP mode the correct system must be specified
in the specifications menu.
The following systems currently can be used:
Infant Flow (Infant Flow or Inspire)
Infant Flow LP
MediJet
When ventilating a patient using a nasal interface in nCPAP or DuoPAP modes,
the clinician must make a compromise between providing additional flow to
compensate for a leak at the nose, or providing an alarm to warn of a leak,
particularly when the appropriate size prongs are small. For a patient who is
capable of spontaneous breathing without support, leak compensation may be set
to a high setting, in order to provide additional flow to compensate for a leak at the
interface between the prongs and the nose. This will provide a more stable
pressure support in the presence of variable leak (for example, with patient
movement). However, in this scenario, the ventilator may fail to provide an alarm
in the event of decannulation.
For a patient who requires an alarm in the event of decannulation, leak
compensation should be set to a low setting. This will provide less stable pressure
support in the presence of variable leak at the nose, but a reliable alarm for
decannulation.
Function areas and ventilation parameters. Ventilation additives.
If Inspiratory Time is set lower than 1 second; the maximum Pressure can NOT be
reached. Depending on the Tubing Set, nCPAP Generator and Humidifier.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ACUTRONIC fabian Therapy evolution and is the answer not in the manual?

ACUTRONIC fabian Therapy evolution Specifications

General IconGeneral
BrandACUTRONIC
Modelfabian Therapy evolution
CategoryMedical Equipment
LanguageEnglish

Related product manuals