64 Ventilation parameters
fabian Therapy evolution | SW 5.1.x
Ref: 121003.EN / Date 2020-01-22
10 Ventilation parameters
Trigger
The trigger sensitivity can be adjusted from level
1 10.
There are 2 trigger functions:
Flow trigger
Pressure trigger
Backup Rate in duoPAP mode
Upper inspiratory pressure in duoPAP mode.
The lower pressure level is selected in duoPAP
mode via the CPAP parameter.
Continuous Positive Airway Pressure
The ventilator provides continuous flow for
producing airway pressure at the set CPAP level
Manual Inspiratory pressure
Applying a manual breath
Next breath is possible after a block period of 200
ms.
Oxygen concentration
Inspiratory oxygen concentration setting
Oxygen flush
Perform O
2
flush / oxygen spray
(Preoxygenation for max. 2 minutes)
Maximum time manual breath*
Maximum duration allowed for manual breaths.
Adjustable from 2 30s.
*In Settings menu
Duration of O
2
Flush*
Adjustable from OFF; 20 120s.
*In Settings menu