44 Configurations menu
fabian Therapy evolution | SW 5.1.x
Ref: 121003.EN / Date 2020-01-22
6.3 Ventilation parameter settings
O
2
flush duration (Preoxygenation) configuration:
(OFF 30 60 120 sec.)
21 Vol% 21 and 100 Vol.%. In cases where oxygen supply source
does not provide 100% oxygen, this should be set to 21% only to
avoid calibration at false value.
Set either Off, Low, Middle or High level of leak compensation (0;
up to 10, 20, 30%) in nCPAP and DuoPAP modes.
Selection of pressure unit (mbar cmH
2
O).
Selection of the nCPAP system used (Infant Flow (Infant Flow or
Inspire), Infant Flow LP, Medijet
®
).
Maximum time
for Manual breath
Maximum Inspiration time for Manual breath (2 to 30 seconds).
optional: The if it is enabled
in service mode.