fabian Therapy evolution | SW 5.1.x
Ref: 121003.EN / Date 2020-01-22
Option C: 5% H
2
O
2
, Hydrogen Peroxide (EPA REG. NO.: 335-1)
1. Prepare a 5% H
2
O
2
solution using purified water.
Use a clean lint-free cloth dampened with the H
2
O
2
solution
2. Wipe the enclosure completely.
Remove any heavy soiling with additional dampened cloths if necessary.
Note: The surface treated must remain visibly moist for five minutes.
Use more dampened cloths if necessary to keep the surface moist for five minutes.
3. Wipe the enclosure with a clean lint-free cloth dampened with purified water for one minute.
4. Allow to dry in the air.
15.2 Accessories
15.2.1 Flow sensor cable
The flow sensor cable must be prepared after each patient treatment.
WARNING:
The flow sensor cable must NOT be sterilized under any circumstances.
15.2.1.1 Preparation for disinfection
1. Switch off fabian before cleaning and unplug at the mains in order to prevent damage due to
penetration by liquid.
2. Remove the flow sensor from the flow sensor cable.
3. Remove any visible soiling on the flow sensor cable with the cleaning agent using a manual
disinfection option listed in section 15.2.1.2.
15.2.1.2 Manual disinfection options
Option A: CaviWipes Surface Disinfectant Wipes (EPA REG. NO.: 46781-14)