84 ion
fabian Therapy evolution | SW 5.1.x
Ref: 121003.EN / Date 2020-01-22
1. Unfold the CaviWipe and wipe the surface of the enclosure completely.
Note: The surface treated must remain visibly moist for three minutes.
Use more wipes if necessary to keep the surface moist for three minutes.
2. Wipe the cable with a clean lint-free cloth dampened with purified water for one minute.
3. Allow to dry in the air.
Option B: 10% bleach (Clorox EPA REG. NO.: 5813-1)
1. Prepare a 10% bleach solution using purified water.
Use a clean lint-free cloth dampened with the bleach solution.
2. Wipe the cable completely.
Remove any heavy soiling with additional dampened cloths if necessary.
Note: The surface treated must remain visibly moist for five minutes.
Use more dampened cloths if necessary to keep the surface moist for five minutes.
3. Wipe the cable with a clean lint-free cloth dampened with purified water for one minute.
4. Allow to dry in the air.
Option C: 5% H
2
O
2
, Hydrogen Peroxide (EPA REG. NO.: 335-1)
1. Prepare a 5% H
2
O
2
solution using purified water.
Use a clean lint-free cloth dampened with the H2O2 solution
2. Wipe the cable completely.
Remove any heavy soiling with additional dampened cloths if necessary.
Note: The surface treated must remain visibly moist for five minutes.
Use more dampened cloths if necessary to keep the surface moist for five minutes.
3. Wipe the cable with a clean lint-free cloth dampened with purified water for one minute.
4. Allow to dry in the air.
15.2.2 Flow sensor
WARNING:
Conduct hygienic preparation of flow sensor prior to each use.
15.2.2.1 Disassembly
Remove the flow sensor from the flow sensor cable.
15.2.2.2 Preparation for sterilization by automated cleaning
Please follow the steps below:
1. Rinse under running tap cold water (<35°C / 95°F) to remove excess soil.