fabian Therapy evolution | SW 5.1.x
Ref: 121003.EN / Date 2020-01-22
2. Pre
recommendation of 10 mL per Liter using utility (tap) water 20°C / 68°F - 30°C / 86°F.
3. Transfer the flow sensor onto the 4-Level manifold rack accessory (or other appropriate rack
system) contained inside the washer for processing.
4. Select the appropriate cycle as listed below:
Phase
Recirculation Time
(Minutes)
Temperature
Detergent Type and Concentration
(If applicable)
Pre-wash 1
02:00
Cold tap water
N/A
Wash 1
02:00
45°C / 113°F
Tap water
(Set Point)
Neodisher Mediclean Forte 10 ml per
Liter
Neutralization
Wash
02:00
Warm tap water
Neodisher Z 1.0 ml per Liter
Rinse 1
01:00
Warm tap water
N/A
Dry Time
7:00
90°C / 194°F
(Set Point)
N/A
5. After the drying process package immediately for sterilization in packaging suitable for vapor
sterilization. Ensure that packaging stays dry before autoclaving.
Note: Articles must be individually single pouched in a qualified pouch (e.g. Cardinal Health self-
sealing pouch CAT# 92713). Each of the non-critical devices in-scope are thermostable and
have been validated for steam sterilization.
15.2.2.3 Sterilization options
The recommended sterilization method is steam sterilization by autoclave. The following validated
cycles are suitable for sterilization:
Option A:
1. Set the autoclave with the following parameters
4 preconditioning pulses
Steam cycle at 134°C / 273°F for 3 minutes
Dry cycle for 30 minutes
2. Individually single pouch the cassette articles
3. Autoclave the cassette articles using the cycle parameters in Step 1
Option B:
1. Set the autoclave with the following parameters
4 preconditioning pulses
Steam cycle at 134°C / 273°F for 4 minutes
Dry cycle for 30 minutes
2. Individually single pouch the cassette articles
3. Autoclave the cassette articles using the cycle parameters in Step 1
Option C:
1. Set the autoclave with the following parameters:
4 preconditioning pulses
Steam cycle at 134°C / 273°F for 5 minutes
Dry cycle for 30 minutes