fabian Therapy evolution | SW 5.1.x
Ref: 121003.EN / Date 2020-01-22
2. Individually single pouch the cassette articles
3. Autoclave the cassette articles using the cycle parameters in Step 1
15.2.2.4 Duration of use
The flow sensor must be replaced after 30 sterilization cycles and also in case that any detected defect is
affecting the accessory performance.