93
Troubleshooting & Service
isopropyl alcohol or cleaning solution for protein removal, such as
Thermo Scientific Orion Cleaning Solution 900021.
4. Sample Probe
The sample probe must be inspected daily to ensure that it is clean
and free of any sample buildup.
Sample Probe
• If the sample probe shows signs of fouling, it must be cleaned by
wiping it with a lint-free cloth dampened with isopropyl alcohol
or cleaning solution for protein removal, such as Thermo
Scientific Orion Cleaning Solution 900021.
Weekly Maintenance Routines:
1. Syringe, Tubing and Pipette
Thoroughly clean the fluid path (syringe, fluid tubing, and pipette)
as outlined below to prevent growth of microorganisms in the liquid
system.
Cleaning Solution (1% sodium hypochlorite)
• Make a solution of 1% sodium hypochlorite by diluting commer-
cial bleach with water or using a commercially available product,
such as Bleach-Rite
®
BRSPRAY16.
*Bleach-Rite
®
is a registered trademark of Current Technologies, Inc.