4. Mount the replacement System Exhaust
Fan to the Chassis with the four screws
and hex nuts. The new fan must be set in
the same position as the one that is being
replaced; the arrow that is molded onto
the body of the fan must be facing the
exterior of the instrument.
5. Route the System Exhaust Fan’s wire
leads through the two Cable Clamps.
6. Connect the A2 Connector on the System
Exhaust Fan to the A2 Connector on the
Cooling System Harness.
7. Reverse steps one through four of the
General Disassembly Instructions to re-
assemble the instrument.
112108RPM Rev0
(OsmoPRO Service Manual)