Fuse Replacement
70011R
Reference: Use this instruction with replace-
ment part 70011R.
Tools Required: Small flat screwdriver
Static Grounding Wrist
Strap.
Warning - Hazardous Voltage
Warning – Internal components may
be damaged by static electricity
CAUTION: Improper connections
may cause damage to the instrument.
CAUTION: A discharge of static elec-
tricity from contact with the human
body or other conductor may damage
system boards or static sensitive
devices. NEVER UNPACK, TOUCH
OR HANDLE ANY PCB WITHOUT
WEARING A GROUNDING
(EARTHING) STRAP TO MINI-
MIZE YOUR STATIC DIS-
CHARGE.
Instruction:
1. Turn off the power, unplug the instrument
and remove the power cord from the
instrument.
2. Use a small flat screwdriver to open the
Fuse Holder Door located on the Power
Entry Module (
Figure 1 & Figure 1B).
70011RPM Rev0
(OsmoPRO Service Manual)
Page 1 of 2
For additional information or technical assistance,
please contact Advanced Instruments Hot-Line
TM
Service Center (U.S. 1-800-225-4034, outside North
America +US 1-781-320-9000).
Figure 1A
Figure 1B
Power
Entry
Module
Fuse
Holder
Door

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Advanced Instruments OsmoPRO and is the answer not in the manual?

Advanced Instruments OsmoPRO Specifications

General IconGeneral
BrandAdvanced Instruments
ModelOsmoPRO
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals