5. Remove the Cable Clamp and hex nut
from the side of the Power Supply &
Controls Assembly (
Figure 3).
6. Unplug the three USB Cables and the
Ethernet Cable, from the Ports located on
the Power Supply & Controls Assembly
(
Figure 4)
7. Rest the instrument on its back side.
Remove the two screws that secure the
Power Supply & Controls Assembly to
the bottom of the Chassis (
Figure 5).
8. Place the instrument in the upright posi-
tion. Remove the two screws that secure
the Power Supply & Controls Assembly
to the rear of the Chassis (Figure 6).
9. Slide the Power Supply & Controls
Assembly out of the Base Assembly.
10. Remove the six screws that mount the
Control PCB to the Power Supply &
Controls Assembly.
11. Carefully, lift up the Control PCB and
unplug the C12 Connector from the bot-
tom of it (
Figure 7).
Figure 3
Figure 4
Figure 5
Figure 6
707011RPM Rev0
(OsmoPRO Service Manual)
Page 2 of 4
Two Screws
USB Ports
Ethernet
Port
Two Screws
Cable
Clamp