OsmoPRO Preventive Maintenance Checklist
Preparations
□ Power off the instrument
□ Dampen a soft lab cloth with a mixture of warm water and soap (Isopropyl alcohol also can be
used to further decontaminate the instrument).
General Exterior Wipe Down Procedure
Wipe down the following surfaces:
□ Touch screen
□ Turntable
□ Front, sides and rear enclosure
OsmoPRO Osmometer Cleaning Procedure
□ Clean the Sample Probe by wiping it with a lint-free cloth dampened with isopropyl alcohol or a
cleaning solution for protein removal, such as Thermo Scientific Orion Cleaning Solution.
□ Clean the Cooling Well using a flexible swab dampened with isopropyl alcohol from the Swab
Cleaner Kit. Insert the flexible swab into the round opening in the Cooling Well.
□ If the exhaust fans on the rear of the instrument are unobstructed by dust or debris, then power
off the instrument and use compressed air to remove the debris.
□ Remove solenoid plunger and spring. If necessary, clean the Solenoid by wiping it with a lint-
free cloth dampened with isopropyl alcohol or a cleaning solution for protein removal, such as
Thermo Scientific Orion Cleaning Solution.