32
The OsmoPRO
®
Micro-Osmometer Service Manual
•Hot-Line™ Technical Service
For any questions or problems regarding the
proper operation of the instrument, please
contact our Technical Services Department:
• Telephone:
• 800-225-4034 (toll-free within the United
States and Canada)
• +1 781-320-9000 (outside of the United
States and Canada)
Fax:
• +1 781-320-3669
OsmoPRO Software Version 1.4
Release Notes
Release 1.4
This release includes updates to help ensure
consistent results within the system specifica-
tions that are defined in our user’s manual:
Accuracy (salt solutions)
0 – 400 mOsm/kg H2O: Mean value ± 3
mOsm/kg H2O from nominal value
400 – 2000 mOsm/kg H2O: Mean value ±
0.75% from nominal value
Release 1.3
This release contains updates to improve the
handling of signals from table and tube prox-
imity sensors to eliminate false errors.
Release 1.2
This release provides updates to adjust the
tube detection threshold to prevent detecting a
tube when no tube is present.
Release 1.1
This release includes updates to implement
limits for standard deviation checking of con-
trols for calibration. It also includes a fix to
the issue when printing some foreign lan-
guage characters and the printer door is
opened.