4. Mount the replacement Cooling Fan, the
Finger Guard and the Air-Flow Diverter
to the Cooling Assembly’s Heatsink with
the four screws and washers. The
replacement fan must be set in the same
position as the old fan; the arrow that is
molded onto the body of the fan must be
facing the Heatsink. In addition, the Air-
Flow Diverter must be equally spaced on
the Heatsink (
Figure 3).
5. Manually raise the Cooling Assembly so
that it is approximately ½” below the
underside of the Stripper Bracket. Secure
the Cooling Fan’s wire leads using the
three Cable Clamps shown in Figure 1.
Be sure not to place any stress on the
Cooling Fan’s wire leads as it is secured
in the first Cable Clamp. A small service
loop is required (
Figure 4).
6. Manually lower the Cooling Assembly.
7. Reverse steps one through twelve of the
General Disassembly Instructions to re-
assemble the instrument.
Figure 3
Figure 4
112094RPM Rev0
(OsmoPRO Service Manual)
Page 2 of 2
Air Flow
Diverter
Service Loop

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Advanced Instruments OsmoPRO and is the answer not in the manual?

Advanced Instruments OsmoPRO Specifications

General IconGeneral
BrandAdvanced Instruments
ModelOsmoPRO
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals