GPIB OPERATION
3-2-26
LF CONTROL
LF
[not used alone]
:ON Turn LF output ON
:OFF Turn LF output OFF
:GEN Select LF Generator
Data type :
None
Allowed suffices : None
Default suffix : None
:MON
Select source monitor mode
Data type :
Character Program Data (any one of: AM1S, AM2S, AMD,
ANG1S, ANG2S, ANGD or OFF, where AM represents
Amplitude Modulation, ANG represents Angular Modulation,
the suffix 'S' indicates Source and 'D' indicates Drive)
Allowed suffices : None
Default suffix : None
Examples:
LF:MON FM;ON
LF:MON ANGD
LF:GEN
LF?
Prepares message containing information on the LF operation
in one of the following formats depending on which LF mode
is currently in use:
:LF:GEN;<status>
:LF:MON <source>;<status>
where: <source> is character program data representing the
source being monitored and <status> is a progra
m
mnemonic indicating whether the output is ON o
r
OFF.
Examples:
:LF:GEN;ON
:LF:MON AM1S;OFF