Agilent 1200 Series Autosampler Reference Manual 121
Repairing the Autosampler 4
3 Remove the stator head, stator face and stator ring. 4 Remove the rotor seal and isolation seal.
5 Install the new rotor seal and isolation seal. Ensure the
metal spring inside the isolation seal faces towards the
valve body.
6 Install the stator ring with the short of the two pins facing
towards you at the 12 O’Clock position. Ensure the ring
sits flat on the valve body.