Agilent 1200 Series Autosampler Reference Manual 121
Repairing the Autosampler 4
3 Remove the stator head, stator face and stator ring. 4 Remove the rotor seal and isolation seal.
5 Install the new rotor seal and isolation seal. Ensure the
metal spring inside the isolation seal faces towards the
valve body.
6 Install the stator ring with the short of the two pins facing
towards you at the 12 O’Clock position. Ensure the ring
sits flat on the valve body.
CellFrame CellFrame
CellFrame
CellFrame