158 Agilent 1200 Series Autosampler Reference Manual
4 Repairing the Autosampler
3 Slide ASM board out of the autosampler. 4 Install new board. Ensure ribbon cables are positioned in
the slot in the board.
5 Reconnect the connectors to the board. 6 Replace GPIB and RS232 connector screws.
On completion of this procedure:
On the new board check the switch setting of
address switch S1, see Ta ble 5 0 on page 252, or
Tabl e 5 1 on page 253.
Note:
An incorrect switch setting (e.g., TEST/ BOOT) may
cause the autosampler to turn in a basic mode
(yellow or red flashing status light). In such a case
turn off the pump, re-set the address switches, and
turn on the pump again.
CellFrame
CellFrame
CellFrame
GPIB
screws
RS232
screws