Agilent 1200 Series Autosampler Reference Manual 159
Repairing the Autosampler 4
4 Install the transport assembly (page 139).
5 Install the top cover and foam (page 134).
6 Turn ON the autosampler.
7 Enter the 10-character autosampler serial number. The serial number can 
be entered using either the control module or the ChemStation, see 
“Entering the Serial Number using the Control Module” on page 163 or 
see“Entering the Serial Number using the ChemStation” on page 165. 
8 Check the firmware revision of the autosampler. If the firmware revision is 
older than the current firmware revision of the autosampler, update the 
firmware, see “Replacing the Autosampler Firmware” on page 166