164 Agilent 1200 Series Autosampler Reference Manual
4 Repairing the Autosampler
autosampler. When the 10-character serial number is entered, press Enter
to highlight the complete serial number. Then, press Done (F6).
6 Turn the autosampler off, then on again. The Records screen should display
the correct serial number for this module.
7 If a ChemStation is also connected, re-boot the ChemStation now as well.
NOTE
For firmware revisions below A02.00 it is very important never to press Done if the Serial#
box is blank. In this case, the module can no longer be recognized by either the control
module or the ChemStation. The main board must then be replaced.