GC Not Ready Symptoms 4
Agilent 7890 Series Troubleshooting 103
The brass knurled nut that holds the FID castle assembly
in place must be tight.
Perform FID maintenance if these parts are corroded or
oxidized.
T-20 Torx screws (3)
Ignitor
Knurled nut