Chromatographic Symptoms 3
Agilent 7890 Series Troubleshooting 51
Poor Peak Resolution
Set column flow to optimum linear velocity.
Install and use deactivated consumable parts in the inlet (for
example, a liner).
Perform column maintenance: Bake out contaminants,
remove the contaminated length of column near the inlet,
and reverse and bake out the column as needed.
Check column installation at both ends.
Select a higher resolution column.