18 Agilent 7890 Series Troubleshooting
2 ALS and Detector Symptoms
Plunger Errors
If the ALS reports a front or back plunger error, check the
following possible causes:
The syringe plunger is sticking or is not securely connected
to the plunger carrier.
The plunger solenoid is binding.
The plunger carrier encoder is inoperable.
The autoinjector plunger carrier mechanism will not move.
The plunger does not move freely due to sample residue or
wear. Install a new syringe, making sure to prime the syringe
with solvent before installing.
Procedure
1 Remove the syringe and check it for plunger stickiness or
binding. Replace the syringe if necessary (7693A, 7650A,
7683).
2 Check the viscosity of the sample against the viscosity
parameter. Reset the viscosity parameter if necessary.
3 Restart the sequence.
4 If the error occurs again, obtain Agilent service.