EasyManuals Logo

Agilent Technologies 7890 Series Troubleshooting

Agilent Technologies 7890 Series
208 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #95 background imageLoading...
Page #95 background image
GC Not Ready Symptoms 4
Agilent 7890 Series Troubleshooting 95
Cannot Set a Flow or Pressure
If you cannot set a flow or pressure using the split/splitless,
MMI, PTV, VI, or cool on-column inlets, do the following:
Check the column mode.
Check that a capillary column is configured to the correct
inlet.
Check the configured column dimensions.
Check that the flow is turned on.
If you cannot set a flow or pressure using the purged packed
inlet, do the following:
Check the column mode.
Check that the flow is turned on.
Check the inlet mode. When using inlet flow control, you can
set only flow control modes for columns. When using inlet
pressure control, you can set the column for both flow and
pressure modes.

Table of Contents

Other manuals for Agilent Technologies 7890 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890 Series and is the answer not in the manual?

Agilent Technologies 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7890 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals