EasyManua.ls Logo

Agilent Technologies 7890 Series - FID Baseline Output at Maximum (8 Million)

Agilent Technologies 7890 Series
208 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chromatographic Symptoms 3
Agilent 7890 Series Troubleshooting 73
FID Baseline Output at Maximum (~8 Million)
If the FID output seems to be stuck at a very high value (up to 8
million counts), check for a shorted collector.
1 Check if the interconnect spring has been bent. Remove the
collector assembly and visually inspect the spring.
2 Disassemble the collector assembly and visually check for
rust buildup on any parts. Replace parts as needed. To avoid
this problem, operate the detector at >300 °C.
3 Check for carbonization in the detector due to injection or
aromatic or chlorinated solvents. To avoid this problem,
operate the detector at >300 °C. Reassemble and install the
collector and operate the detector using higher air and
hydrogen flows (air must be 450 mL/min, hydrogen at 35
mL/min).

Table of Contents

Other manuals for Agilent Technologies 7890 Series

Related product manuals