GC Not Ready Symptoms 4
Agilent 7890 Series Troubleshooting 109
Blinking Not Ready Light: Detector Hardware Fault/TCD Filament Voltage
If the TCD filament is open, the Not Ready LED blinks and the
GC will not become Ready.
1 Press [Status]. If the display reads Front detector hardware fault
or (Back or Aux detector hardware fault), the problem may be
the TCD filament.
2 Press [Front Det] (or [Back Det] or [MS/Aux Det]).
3 For GCs with firmware B.02.02 or higher, the display will
read either TCD Open Filament for an open filament, or TCD
Bridge Voltage for a shorted filament.
For GCs with earlier firmware: If the display reads Fault: TCD
Filament Voltage, scroll to the Filament setpoint. If the setpoint
is flashing Off, the problem is an open TCD filament.
4 Power cycle the GC.
5 Press [Front Det] (or [Back Det] or [MS/Aux Det]).
6 Scroll to the Filament and press [Off/No] to disable the TCD
filament. Contact Agilent for service.
To continue to use the GC without the TCD:
1 Turn off all TCD parameters. Save the method.
2 Reconfigure the GC to use another detector. Move the column
and set other parameters as needed.

Table of Contents

Other manuals for Agilent Technologies 7890 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890 Series and is the answer not in the manual?

Agilent Technologies 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7890 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals