EasyManuals Logo

Agilent Technologies 7890 Series Troubleshooting

Agilent Technologies 7890 Series
208 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #80 background imageLoading...
Page #80 background image
80 Agilent 7890 Series Troubleshooting
3 Chromatographic Symptoms
NPD Solvent Quenching
If the baseline does not recover after a solvent peak, try the
following:
Turn hydrogen off/on around the solvent peak.
Use nitrogen as the makeup gas.
Set the total column flow and makeup gas to less than
10 mL/min.
Increase the air flow by 10 mL/min.
Increase the detector temperature to 325 °C.
Implement an Agilent Dean’s switch solvent vent solution.

Table of Contents

Other manuals for Agilent Technologies 7890 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890 Series and is the answer not in the manual?

Agilent Technologies 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7890 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals