86 Agilent 7890 Series Troubleshooting
3 Chromatographic Symptoms
Negative Peaks Seen with TCD
Verify that the correct gas type is being used.
Check for a leak in the system, especially at the detector
column fitting. (See “Checking for Leaks”.)
Consider thermal conductivity of analytes relative to carrier.
Check the flow settings, then measure the actual detector
flows. (SeeTo Measure a Detector Flow”.)