Chromatographic Symptoms 3
Agilent 7890 Series Troubleshooting 87
TCD Baseline Has Dampened Sinusoidal Noise Trailing Peaks (Ringing
Baseline)
Wrong data rate is selected in the data system. For TCD, the
data rate should be
5 Hz.