Troubleshooting Tasks 8
Agilent 7890 Series Troubleshooting 197
To Measure FID Leakage Current
1 Load the analytical method.
Make sure flows are acceptable for ignition.
Heat the detector to operating temperature or to 300 °C.
2 Turn off the FID flame.
3 Press [Front Det] or [Back Det], then scroll to Output.
4 Verify that the output is stable and < 1.0 pA.
If the output is unstable or > 1.0 pA, turn off the GC and
check for proper assembly of the upper FID parts and for
contamination. If the contamination is confined to the
detector, bakeout the FID.
5 Turn on the flame.