GC Not Ready Symptoms 4
Agilent 7890 Series Troubleshooting 91
Flow Never Becomes Ready
If the gas flow never becomes ready, check for the following:
Check the supply gas for sufficient delivery pressure.
Check the restrictors installed in the Aux EPC module. See
Restrictors in the Advanced User Guide.
Check the configured gas type. The configured gas type must
match the actual gas plumbed to the GC.
Check for leaks in the gas delivery plumbing and the GC. (See
Checking for Leaks”.)