EasyManuals Logo

Agilent Technologies 7890 Series Troubleshooting

Agilent Technologies 7890 Series
208 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #116 background imageLoading...
Page #116 background image
116 Agilent 7890 Series Troubleshooting
5 Shutdown Symptoms
Detector Air Shutdowns
When air is used to combust a fuel gas (hydrogen) in a detector,
an insufficient air supply will cause a shutdown for the air
channel. In addition, the Air and Hydrogen flows for the
associated detector will be turned Off.
To recover from this state.
1 Fix the cause of the shutdown. Verify the air supply. The FID
and NPD require a delivered gas pressure of 380–690 kPa
(55–100 psi). The FPD and FPD
+
require a delivered gas
pressure of 690–827 kPa (100–120 psi). See the Agilent GC,
GC/MS, and ALS Site Preparation Guide.
2 Press the key for the detector that initiated the shutdown.
Scroll to each pneumatic parameter that is flashing Off, then
press [On] or [Off].

Table of Contents

Other manuals for Agilent Technologies 7890 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890 Series and is the answer not in the manual?

Agilent Technologies 7890 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7890 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals